The Introduction of Sterile Barrier System

Medical packaging materials are also known as "SBS" Sterile Barrier System, or "sterile barrier system."
 
basic requirements
1. The formation of a sterile barrier system requires the following requirements:
2. The sterilization medium should be able to effectively penetrate the medical packaging materials.
3. In the sterile barrier system, the sterilization medium can be effectively sterilized.
4. Medical packaging materials should be able to protect the surgical instruments inside the package.
5. Medical packaging materials should be able to discharge the sterilization medium.
6. The sterile barrier system formed by medical packaging materials should be able to ensure its sterility. From the time of sterilization to the time of opening, the inside should always be sterile.
Related standards
1.ISO 11607-1, 11607-2
2.EN 868 series standard
3.ASTM-various experimental methods
4.AAMI ST77
The above standards are currently the most commonly used standards for sterile barrier systems.
 
Authentication method
Two tests are required to validate a sterile barrier system:
1. Physical tests to demonstrate the stability of the material and the effectiveness of the sterilization period.
2. Microbial barrier test to demonstrate the effectiveness of the packaging material for bacteria inhibition.
For packaging materials such as paper-plastic bags, it is necessary to check the seal integrity and seal strength. Detection methods include: sealing performance dye test, visual inspection and peeling test.
The most important indicator of microbial barrier experiments is BFE (Bacterial Filtration Efficiency), which is the efficiency of bacterial filtration.
For non-woven bags and rigid containers, because they do not use sealed packaging, there is no need for seal integrity testing, and their sterile barrier system testing requires the use of microbial aerogels for testing the entire sterilization package.
Remarks: EN ISO 11607-1:2014 has new changes to the definition of microbial barriers, but the basic performance requirements for the sterile barrier system to be transported, dispensed, stored, etc. have not changed.
 
2. Types of sterile barrier systems
The sterile barrier system is mainly divided into three categories.
1. A disposable sterile barrier system that achieves a sterile barrier function through a sealed package. Such as paper plastic bags.
2. A disposable or reusable packaging material that achieves a sterile barrier function through a closed package. Such as textiles.
3. A reusable microbial barrier system that achieves a sterile barrier function through a valve or filter element. Such as a rigid container.
For flexible packaging, the microorganisms are blocked by the bypass path formed during the packaging process. The rigid container is subjected to microbial barrier by its own valve structure, labyrinth structure and the like.
All three sterile barrier systems have strict testing standards to detect their materials and bacteriostatic effects.
 
3. Aseptic barrier system test method
Physical experiments on disposable packaging materials include: material weight, tensile strength, and the function of its microbial barrier, and stripping experiments are required to test seal strength. Different experiments are required depending on the material.
The experiment was designed based on the material of the sterile barrier system and the principle of inhibition of bacteria.
For reusable rigid containers, the above requirements also need to be met. Since they are not sealed as a sterile barrier system, there is no need to test the sealing performance, but the bacteriostatic structure needs to be tested.

评论

此博客中的热门博文

Reference Standard For Seal Verification of Sterilization Pouches For Medical Devices

With the increasing demand for  sterilization pouches  of medical devices, quality and safety inspection have become more and more stringent. sterilization pouches of medical devices (blister box +  tyvek roll  material,  sterilization paper bag , etc.) as a form of packaging directly in contact with the device and used to block microorganisms. No matter what form of packaging, it has two characteristics, which can meet the sterilization operation and remain sterile until the transportation conditions are used. Therefore, relevant departments began to focus on the review requirements of sterile packaging and make specific requirements for its performance and quality.   As we all know, GB/ t19633-2005 "packaging of ultimately sterilized medical devices" is one of the authoritative standards in the current sterile packaging standards. It is equivalent to iso11607-2003 "packaging of ultimately sterilized medical devices". Clarify the basic performance req...

Medical Sealing Machine Operation & Maintenance Manual(I)

  I. About Thank you for choosing  KM100 series of sealing machine . To maximize its performance,please carefully read this manual and install and operate it by strictly following this manual to ensure its safety and reliability in use. We recommend placing the manual in an accessible place for better reference during the operation. II. Applied Area Automatic  medical sealing machine , 100-L series, is capable of continuous paper-plastic bag, 3D paper-plastic bag and paper-paper bag sealing. The sealing process satisfies the requirements of high-temperature steam sterilization, low-temperature ethylene oxide sterilization, hydrogen peroxide plasma sterilization and radiation sterilization. Seal quality meets corresponding international standards and GMP recognition. Sealable materials: Compliance with EN868-5 and YY/T 0698-5 bags and volumes; Compliance with EN868-4 and YY/T0698-4 paper bags; High density polyethylene materials (such as Tyvek); Aluminum foiled materials. ...

Definition And Use Of Sterilization Or Chemical Indicator

  Sterilization indicator  tape and label   Definition: Wrinkle paper or label paper as the base material, coated with color-changing ink as sterilization indicator, back coated with pressure sensitive adhesive.  The indicating tape used for chemical monitoring outside the pressure steam  sterilization package , when reaching certain sterilization conditions 121℃, 20 minutes or 132℃, 3 minutes, the chemical indicating stripe on the back base of wrinkled paper will change from light yellow to black.   USES:   1. Used to mark whether the package has been treated by pressure steam sterilization process.   2. If the discoloration is uneven or incomplete, it may indicate that the package has not been sterilized in accordance with the conditions.   3, wrap the parcel (if no sealing tape is used).   4, sterilization information can be recorded (operator, sterilization period, item name, pot, etc.).   Chemical instruction card for...